“COMPACT” ULTRASOUND NEBULIZER Model WN-116 U Instruction Manual Table of contents 1. Introduction 2. Recommendations 3. Spare Parts Description 4. Use Instructions 5. Operation with Car Lighter Adapter (Optional Accessory) 6. Operation with rechargeable battery pack (Optional Accessory) 7. Cleaning and Maintenance 8. Air Filter Change 9. Optional Accessories 10. Trouble Shooting Chart 11. Technical Specifications 12. Symbols Key 13. Certifications 14. Warranty 2 2 3 4 5 5 7 7 8 9 10 10 11 11 1 Introduction Congratulations for having purchased a B.Well “Compact” Ultrasonic Nebulizer! Thanks for your support.B.Well warrants that this device has been produced using high quality materials and it complies with the safety rules in force both domestically and internationally.The B.Well compact ultrasonic nebulizer operates on ultrasound which produces the drug solution vibration and turns it into mist, these tiny particles get deposited in the lungs and consequently, effective and quick results are achieved and no side effects are caused. This model comes with the exclusive Closed Water Chamber System, easy to use: all you need to do is to pour the solution into the medication cup and it is ready to use. It operates on two kinds of power supply: mains supply and car lighter adapter (optional accessory). 2. Recommendations 2 1. Read the manual carefully before using the device and keep it in case you need further information in the future. 2. Do not turn the device on if there is no drug solution into the medication cup. 3. Disconnect the nebulizer when it is not being used. 4. Do not remove the medication cup cover while the device is operating. 5. Do not expose the nebulizer to a heat source. 6. Do not pour more than 8 ml of solution into the medication cup. 7. Do not leave the drug solution into the medication cup when the device is not being used. 8. Do not scrape or pierce the medication cup cover. 9. Do not touch the medication cup or the drug solution while the device is being used. 10. Do not dip the nebulizer in water. 11. Children may only use the device under adult supervision. 12. To avoid infections and contagion the accessories should not be shared by different patients. They are intended for personal use. 13. Use the unit intermittently. Maximum operating period: 30 minutes. Rest period: 30 minutes. 14. B.Well warrants the quality of the liquid contained in the Closed Water Chamber for a period of 3 years. This solution can be changed easily by the authorized after sales agents. 15. This device can only be repaired by B.Well Customer Service Center. 3. Spare Parts Description Top cover Medication cup Power Indicator Adapter Main body Clear Housing Air Filter ON/OFF Switch Power Connector Air Chamber 3 AC Adapter Car lighter adapter (optional) Nasal piece Mouth piece 4. Use Instructions 1. Take the unit and the AC adapter out of the box. 2. Remove the clear housing and pour the medication into the medication container (8 ml max.). 3. Put the clear housing and insert the device to be used into it (adult mask, pediatric mask, and nasal or mouth piece). CAUTION: carefully check the membrane placed at the bottom of the medication cup before each use. If it is pierced or broken do not use the nebulizer and contact the B.Well Customer Service Center. 4. Plug the unit and the power connector to the unit. 5. Press ON switch, the green light will turn on and the unit will start to operate. Adjust the air regulator to the desired intensity. 6. The unit will automatically turn off at 3 minutes. If you wish to continue nebulizing, check if the medication amount in the pool is the suitable one. Add more if necessary. Then press the ON/OFF switch twice and the unit will start to operate again. 7. Once use is over, unplug the equipment and disconnect the AC adapter. Remove the clear housing and dry the left medication using a clean cloth. 4 5. Operation with Car Lighter Adapter (Optional Accessory) Connect the adapter to the car lighter and the other wire end to the unit. Then operate the nebulizer as per the instructions previously described. 6. Operation with rechargeable battery pack (Optional Accessory) Precautions and Safety Measures - Do not use the device under the sun, or close to a heat source. - Never immerse the battery pack in water. - Do not operate the unit for more than 20 minutes. - It is normal for the Pack to increase temperature while being operated. - If the Pack has not been used for a long period (3 months or more), repeat the initial charge. - If the Pack will not be used for a long period of time, it is advisable to store it fully charged. Battery Charging / Initial operation The Battery Pack is initially discharged and it is necessary to fully charge it before its first use to ensure the capacity and useful life of the battery. A.1 - Make sure that the ON/OFF button is OFF (0). A.2 - Connect the supply power cord of your nebulizer to the mains socket. Connect the plug of the power supply cord into the opening of the pack base and charge 14 hours. Battery Charging / Regular operation A.1 - Make sure that the ON/OFF button is OFF (0). A.2 - Connect the supply power cord of your nebulizer to the mains socket. Connect the plug of the power supply cord into the opening of the pack base and charge between 6 and 12 hours, depending on the use. Do not leave the Pack charging for more than 24 hours. 5 Operating Instructions Prepare the nebulizer as you normally do or follow the instructions. Connect the pack plug to the nebulizer. Turn the ON/OFF button to the ON position (I) (the green light that indicates the state of the battery will illuminate). - Turn the nebulizer on. If the red light is on disconnect the nebulizer plug and charge the battery pack. Battery autonomy at full charge is approximately 30 minutes. Once nebulizer use has been completed, turn the Battery Pack off and disconnect it. Technical Specifications Battery Type Capacity Output Input Charge time 6 nickel-metal hydride 550mAH 12V --- 1.1A 12V --- 0.5A 14 hours Initial Charge 6 to 12 hours Normal Charge 7. Cleaning and Maintenance Important: Clean the accessories before using the nebulizer for the first time or after each use following these instructions: wash all the pieces, except for the air filter, with warm water and detergent. Rinse with warm tap water to remove the detergent. Dip the pieces into a clean bowl containing 3 parts of warm water and 1 part of vinegar during 30 minutes, disinfectant can also be used. Wash your hands and take out the pieces from the disinfectant solution, rinse them with warm tap water and leave them on an absorbent piece of paper. Keep the pieces into a hermetic bag. Do not dry them with a towel. Clean the outer surface with a dry piece of cloth. WARNING: In order to avoid infections, it is highly recommended to clean the pieces after each use. Only nebulize solutions prescribed by a doctor and follow his instructions carefully. Do not use balm substances. 8. Air Filter Change 1. The air filter has to be replaced if it turns dirty or torn. 2. Remove the air chamber from the medication cup cover. 3. Take out the filter and place a new one. 4. Put all the pieces back. 7 9. Optional Accessories Illustration 8 Description Quantity, items Code Number Adapter 1 NA-P116U Clear housing 1 NA-L116U Air filter 3 NA-F116U Pediatric mask 1 NA-MC116U Adult mask 1 NA-M116U Mouthpiece 1 NA-N116U Nasal piece 1 NA-V116U Rechargeable battery (optional) 1 NA-B114K AC Adapter 1 HK-A215-A12 Car adapter (optional) Instructions manual and warranty 1 NA-S116U 1 10. Trouble Shooting Chart If something wrong happens while the device is working, please check connected and correct the following points: Symptom Possible cause Suggested solutions Although the device is on, it does not work 1. Incorrectly plugged 2. No power 3. The AC Adapter is not connected 1. Check the plug is properly 2. Check the power is not gone out 3. Check the AC Adapter is properly connected The mist does not come out There is no solution into the cup Pour some drug solution into the medication cup Little mist is delivered 1. The air filter is dirty 2. There is no enough solution into the cup 3. Little power 4. Air excess 1. Change the filter 2. Pour some drug solution into the medication cup 3. Increase the power 4. Regulate the air flow The device operates for a period and then turns off Timer off Press the key twice and turn it on again 9 11. Technical Specifications Method of nebulization ultrasonic vibration Operating voltage Pri: 100 - 240V, 50 - 60Hz, 0.45A Sec: 12Vdc, 1.25A Class Power 16 W Oscillation frequency 2.5 Mhz Nebulization rate 0.5 ml/min Particle dimension 1.5 – 5.7 (MMAD 3.8 microns) Medication cup maximum capacity 8 ml Weight 260 g Device dimensions 83.5 x 169 x 60 mm Operating temperature and humidity 5° to +40°C, 30% to 95 % RH Storage temperature and humidity -25° to +60°C, 10% to 95% RH Atmospheric pressure range for transport and storage 500-1060 mb * The specifications and the outer design are subject to change for improvement reasons without previous notice. 12. Symbols Key On ................................... Off ......... ......................... AC ...................... ............ Class II ........................ Type B ..... ........................ Read the instructions .......... 10 13. Certifications B.Well meets the highest international quality standards and was awarded the following certificates: ISO 9001 Quality Systems (International) EN 46001 Medical Equipment (European Community) 93/ 42/ CE Directive (European Community) 510 (k) (USA) (Australia -New Zealand) (Argentina) (Argentina ) Manufactured by: B.Well Limited 758 Great Cambridge Road, Enfield, Middlesex, EN1 3PN, United Kingdom 14. Warranty B.Well warrants to the original retail purchaser that this product is free from defects in material and workmanship for a period of 3 years from the date of purchase. This warranty does not cover failures resulting from misuse, abuse, alteration of the B.Well Nebulizer System, use improper components or failure to comply with the operating instructions. This warranty does not include the Nebulizer System tubing or accessories. 11 ULTRASONIC INHALER B.Well Model WN-116 U Operating instructions Contents 1. Introduction 2. Recommendations 3. Main parts 4. Instructions for use 5. Using the car adapter (supplied separately) 6. Working with the battery (supplied separately) 7 Processing and maintenance 8. Replacing the air filter 9. Components 10. Table of possible faults 11. Technical characteristics 12. Symbols 13. Certificates 14. Warranty 14 14 16 17 17 18 20 20 21 22 23 23 24 24 1. Introduction Dear buyer! Congratulations on your purchase of the B.Well compact ultrasonic nebulizer! Thank you for your support. B.Well guarantees that this device is manufactured from high quality materials and complies with current national and international safety standards. The action of the compact inhaler WN-116 U is based on the use of ultrasound, which causes vibration medicinal solution and turns the latter into an aerosol cloud. The smallest particles are delivered directly to the lungs, which causes a rapid therapeutic effect without side effects. This model is equipped with an exclusive “Closed Water Reservoir” system, which greatly simplifies the use of the device: all you need to do is pour the inhalation solution into the medicine container, and the device is ready for use. The device can operate from two power sources: from the mains and from a car adapter (optional accessory). 2. Recommendations 14 1. Before using the device, carefully read the instructions and keep them for the entire service life of the device. 2. Do not turn on the device if the medicinal solution is not poured. 3. Disconnect the device from the power supply if you are not using it. 4. Do not disconnect the aerosol chamber while the device is running. 5. Do not store or use the device near a heat source. 6. Do not pour more than 8 ml of inhalation solution into the container. 7. Do not leave inhalation solution in the medicine container if you are not using the device. 8. Do not damage (scratches, punctures) the transparent membrane or medicine container. 9. Do not touch the medication container or the inhalation solution itself while the device is running. 10. Do not immerse the device in water. 11. Children may only use the device under adult supervision. 12. To avoid the spread of infections, components should not be used by different patients. They are intended for individual use only. 13. Use the device intermittently. Maximum period of continuous use: 30 min., break: 30 min. 14. B.Well company provides a 3-year guarantee on the quality of the liquid contained in the closed water tank. This solution can be easily replaced by specialists from authorized service centers. 15. Repairs to this product should only be carried out by service centers recommended by B.Well. 15 3. Main parts Top cover of the device Medicine container Indicator Adapter Main unit Aerosol chamber Air filter Air chamber AC adapter Nose nozzle Mouthpiece 16 ON/OFF button Adapter socket Car adapter (supplied separately) 4. Instructions for use 1. Open box and remove the inhaler and all accessories. 2. Disconnect the aerosol chamber and pour the inhalation solution (no more than 8 ml) into the medicine container. 3. Reinstall the aerosol chamber. Connect the mask. The nose piece or mouthpiece is connected via an adapter. WARNING: Carefully inspect the diaphragm of the medication container before use. If it is damaged, do not use the device and contact a service center. 4. Connect the device to an AC power supply (220 V, 50 Hz) using a power adapter. 5. Press the ON/OFF button, the indicator will light up and the device will start working. Adjust the aerosol flow from the inhaler using the air flow regulator. 6. The device will automatically turn off after 3 minutes. If you want to continue inhalation, check whether there is a sufficient amount of inhalation solution in the medicine container and add it if necessary. Then press the ON/OFF button twice and the device will start working again. 7. Upon completion medical procedure turn off the device, disconnect the plug of the network adapter, and then the network adapter from the power supply. Disconnect the aerosol chamber and remove any remaining inhalation solution using a clean, soft cloth. 5. Using a car adapter (supplied separately) Connect the car adapter to the car's on-board power supply (12V) and directly to the inhaler. Turn on the device and use according to the general instructions. 17 6. Handling the battery (optional) Precautions - Do not use the battery in direct sunlight or near a heat source. - Do not immerse the battery in water. - Do not use the device for more than 20 minutes. - It is normal for the battery pack to become warm when using the device. - If the battery has not been used long time(3 months or more), it should be fully charged. - If the battery will not be used for a long period of time, it is recommended to store it fully charged. Battery charge. Initial Actions The battery pack is initially completely discharged and must be fully charged before first use to ensure long life. A.1 – Make sure the (On/Off) button is in the (Off) position. A.2 – Connect the adapter of your inhaler to the electrical outlet. Insert the plug of the adapter's electrical cord into the socket on the base of the battery pack and charge for 14 hours. Battery charge. Continuous Operation A.1 – Make sure the (On/Off) button is in the (Off) position. A.2 – Connect the adapter of your inhaler to the electrical outlet. Insert the electrical cord plug into the socket on the base of the battery pack and charge for 4 to 12 hours. The battery charging time should not exceed 24 hours. 18 Instructions for use Prepare the inhaler for use. Connect the plug battery to the inhaler. Move the (On/Off) button to the (On) position. The green indicator on the battery pack will light up - the battery is charged. Turn on the inhaler. If the red light is on, disconnect the inhaler and charge the battery pack. Autonomous operation Battery life when fully charged is 30 minutes. When the inhaler has finished working, turn off the battery pack and disconnect it. Battery Specifications Battery type Nickel-magnesium Capacity 550mAH Output 12V --- 1.1A Input 12V --- 0.5A Charging time 14 hours initial charging, 4-12 hours normal charging 19 7. Handling and maintenance Important: Necessary treat all components before using the device for the first time and after each procedure according to the following instructions: wash all parts except the air filter and adapter, warm water And detergent. Rinse under warm running water to remove detergent. For 30 minutes, immerse all parts in a solution containing 3 parts warm water and 1 part vinegar, or in a bactericidal disinfectant sold in pharmacies. Wash your hands, remove parts of the device from the disinfectant solution, rinse under running water warm water and dry by placing on a paper towel. Store the device parts in a sealed bag. Do not dry with a towel. The outer surface can be wiped with a clean dry cloth. WARNING: To avoid the spread of infections, it is recommended to clean parts of the device after each use. Use only medications prescribed by your doctor for treatment and strictly follow his instructions. Do not use essential substances. 8. Replacing the air filter 1. The air filter must be replaced if it is mechanically damaged or heavily soiled. 2. Disconnect the aerosol chamber and remove the air filter using tweezers. 3. Install a new air filter. 4. Attach the aerosol chamber. 20 9. Accessories Accessories Name Quantity, pieces Code Adapter 1 pc. NA-P116U Aerosol chamber 1 pc. NA-L116U Air filter 3 pcs. NA-F116U Children's mask 1 pc. NA-MC116U Mask for adults 1 pc. NA-M116U Mouthpiece 1 pc. NA-N116U Nose tip 1 pc. NA-V116U Rechargeable battery 1 pc. NA-B114K Network adapter 1 pc. HK-A215-A12 (optional) 21 Car adapter 1 pc. Instruction manual and warranty 1 pc. (sold separately) NA-S116U 10. Table of possible malfunctions If any malfunctions occur, carefully study the table and try to eliminate them. 22 Malfunctions Possible cause Remedy The device does not work. 1. Incorrect connection of the device. 2. There is no voltage in the electrical network. 1. Check the power cord connection. 2. Check the network voltage. 3. Check the network adapter connection. No spraying. There is no medicine in the medicine container. Fill the medicine container (no more than 8 ml). Insufficient aerosol formation. 1. The air filter is dirty. 2. There is no medicine in the medicine container. 3. There is no voltage in the electrical network. 4. Excess air. 1. Replace the air filter. 2. Fill the medicine container. 3. Check the power supply voltage. 4. Adjust the air flow. The device runs for a while and then turns off. The sleep timer has tripped. Press the ON/OFF button. twice and continue the procedure. 11. Technical characteristics Spray method ultrasonic vibration Operating voltage Pri: 100 - 240V, 50 - 60Hz, 0.45A Sec: 12Vdc, 1.25A Class Power 16 W Oscillation frequency 2.5 MHz Spray speed 0.5 ml/min Size particles (microns) 1.5 -5.7 (average aerodynamic size 3.8) Max. medicine cup capacity 8 ml Weight 260 g Device dimensions 83.5 x 169 x 60 mm Operating temperature and humidity from +5° to +40°C, 30% - 95% Storage temperature and humidity from -25° to +60° C, 10% - 95% Atmospheric pressure range for transportation and storage 500 -1060 millibars * Specifications and design of the device are subject to change for improvement without prior notice. 12. Symbols Enabled................... Disabled...... ........ Alternating Current (AC) ........ Class II ........................... .. Type B................... Read the instructions........... 23 13. Certificates The high quality of the device and components is documented. In Russia: Registration certificate of ROSZDRAVNADZOR (FS No. 2006/91 dated January 30, 2006). Certificate of conformity of the State Standard of Russia. Sanitary and epidemiological report of ROSPOTREBNADZOR. International certificates: Manufactured by: B.Well Limited 758 Great Cambridge Road, Enfield, Middlesex, EN1 3PN UK 14. Warranty B.Well guarantees the quality of materials and workmanship for 3 years from the date of purchase of the device. The warranty does not cover damage resulting from misuse, accidents, modifications made to the device by third parties, the use of other components, or as a result of actions contrary to the instructions in the Instructions. The warranty does not apply to masks, mouthpieces and other components and accessories. 24

The compact MESH nebulizer B.Well WN-114, for children, (or electronic mesh, membrane) is intended for the treatment of acute respiratory diseases, chronic diseases respiratory tract, bronchial asthma, pneumonia and other respiratory diseases in medical institutions and at home.

The operating principle of the B.Well WN-114 electronic mesh inhaler is based on innovative technology spraying the medicine using a mesh membrane. In this device for aerosol therapy, atomization of the drug occurs when the flow of the drug passes through the smallest holes in the mesh membrane. An important feature of the technology is that ultrasonic vibrations of low frequencies are not applied to medicinal product, as in traditional compact devices, but onto a grid, as if pushing drug molecules through the microscopic pores of the membrane. With this spray system, the drug is not destroyed, since there is no direct effect on the drug molecules. The technology is also characterized low level noise. The special design of the spray chamber of the MESH nebulizer B.Well WN-114 allows for inhalation at an angle of up to 45° without compromising the spraying process, which makes it convenient for inhalation for bedridden patients and small children during sleep.

The B.Well WN-114 membrane nebulizer can use the entire list of medications recommended for nebulizer therapy. The type of drug, dose and mode of administration can only be prescribed by the attending physician.

Before using the B.Well WN-114 inhaler for the first time, all parts and components of the device must be disinfected. The spray chamber can be disinfected by boiling. Masks, mouthpiece and device body must not be boiled. To disinfect, wash the mouthpiece, adapter and masks under running water and place them in a disinfectant solution (alcohol disinfectant solution or 3% hydrogen peroxide solution) for 10 minutes, then the attachments must be rinsed and completely dried. Before use, the device, its parts and components must be completely dry.

Today, one of the effective and affordable methods of treating diseases affecting the respiratory system is nebulizer therapy. In almost any pharmacy you can see several models of inhalers, varying in design, cost and characteristics. In particular, B. Well brand devices, which provide all the technical features for inhalation therapy. The lines of nebulizers of this brand meet modern global requirements in this area.

    Show all

    Types of B. Well inhalers

    The safety and quality of B. Well devices are confirmed registration certificates Federal Service for Surveillance in Healthcare, declarations of conformity to GOST, declarations and certificates of compliance with technical regulations of the Customs Union. The official representative of B. Well Swiss in Russia is Alfa-Medica.

    Inhalers of this brand can be divided into 4 groups. The first 3 of them are combined under the general name “nebulizer” (from the Latin nebula - cloud, fog). Devices of this type, unlike steam devices, create an aerosol from an inhaled solution.

    MESH nebulizers (electronic mesh)

    Characteristics:

    • compact;
    • light (the weight of the B. Well MESH nebulizer is 137 g);
    • have innovative spraying;
    • suitable for all medications for nebulizer therapy (when nebulized using Mesh technology medicinal substance does not collapse, which allows the use of all medications for nebulizers, including antibiotics, mucolytics and hormonal drugs);
    • silent;
    • work even when tilted up to 45 degrees.
    • particles ≈ 3.4 microns in size;
    • the share of the respirable fraction (the percentage of particles of the required size in the inhaled aerosol) is 60-70%, depending on the characteristics of the drug;
    • the residual volume of the drug is less than 0.15 ml.

    It is possible to disinfect and clean the medicinal chamber by boiling. Allowable boiling time is up to 4 minutes. Power sources used: 2 AA batteries or AC adapter.

    Models of MESH nebulizers B. Well: WN-114 adult , WN-114 child.


    Ultrasonic

    The ultrasonic device is the predecessor of the MESH nebulizer. Particles are formed under the influence of ultrasonic waves high frequency. Due to the fact that under the influence of ultrasound, high-molecular compounds are destroyed, which are active substance medications, the therapeutic effect of therapy can be significantly reduced. For this reason, for use in ultrasonic inhalers Only certain medications are allowed.

    Compressor

    An aerosol cloud is formed using a compressor. When working with them, you can use all types medicines, approved for nebulizer therapy, without loss of their effectiveness. Used to treat both adults and children. They usually have additional accessories: a mask for babies, a nasal shower and others. They are reliable and for a long time operation.

    Features of B. Well nebulizers:

    • the average particle size is ≈ 3 µm;
    • the share of the respirable fraction (the content in the cloud of sprayed vapor of particles whose size is less than 4 microns) is more than 70%;
    • High power compressor, economical consumption of the drug.

    Power source: AC adapter.

    Compressor nebulizers B. Well are represented by models:

    • MED-121.
    • MED-125.
    • PRO-110.
    • PRO-115.
    • WN-117.
    • WN-112.

    Steam inhalers

    Their mechanism of action is based on the evaporation effect hot liquid. Steam inhalations are effective for coughs and runny nose. For inhalations can be used in a steam appliance essential oil, decoction and infusion of herbs, mineral water . The device has a special temperature regime, which allows you to preserve the maximum possible amount of useful substances in the steam. With such inhalation, the possibility of getting a burn is excluded, since the solution is located in a sealed chamber, and the steam temperature is as comfortable as possible. The steam inhaler can also be used for cosmetic procedures.

    Characteristics of B. Well steam inhalers:

    • particles larger than 10 microns;
    • soft steam temperature +43 °C;
    • The kit includes masks for medical and cosmetic procedures.

    Model steam inhaler B.Well: WN-118.


    B. Well inhalers for children

    The B. Well brand also has nebulizer models created specifically for children. For example, a device that looks like a toy train. The package includes colored stickers to complete the look. The B. Well MED-125 inhaler model also has a children's design. The device comes with stickers - animals and hearts. To carry out procedures for children in the first year of life, B. Well inhaler models are equipped with infant masks.


    Instructions for use

    For therapy with a nebulizer, do not use:

    • Substances and solutions that contain a suspension of particles (decoction, suspension, infusion, etc.). They are much larger than particles of the respirable fraction. Their use in nebulizer therapy can be harmful to health.
    • Solutions containing oils, including essential oils. When oil particles enter the lower respiratory tract, they can form tiny films and increase the risk of “oil pneumonia.” An exception is the use of steam inhalers.
    • Flammable anesthetic mixtures that ignite easily on contact with air, oxygen or nitrous oxide.
    • Aromatic substances.

    It is necessary to carefully monitor the expiration date of solutions used for inhalation. Using medications with expired suitability is unacceptable.

    In order to prevent the spread of infections, the use of the same inhaler components by different patients without prior disinfection is unacceptable. Disinfection methods that can be used are indicated in the instructions for each device.

    If components are used for the first time or after a long break, cleaning and disinfection are carried out before use. In the future, it should be carried out every time after using the device, since the remaining drug can harden and cause damage to the mechanism. After cleaning, dried parts must be stored in a clean, dry place out of the reach of children.

    Do not leave cleaning solution in the mouthpiece, chamber or on the mask. Do not use a microwave oven, hair dryer or other household appliances to dry the components of the nebulizer (air hose, diffuser, etc.).

    It is important to follow the following rules:

    • When using the nebulizer, do not cover it with a blanket, scarf or towel.
    • Do not expose the device to harmful vapors or volatile substances.
    • It is not recommended to pour a volume of solution into the medicine container that exceeds the maximum volume for this model.
    • Before using the inhaler, it is important to ensure that there are no malfunctions and/or visible damage. If there is any doubt about its functionality, it should not be used.
    • If a defect and/or abnormal operation is detected, the device must be switched off immediately.
    • The nebulizer must be used exclusively for its intended purpose.
    • The device must not be used to spray medications in the form of powders, acidic medications, or water. The liquid used is saline solution (isotonic sodium chloride solution).
    • There is a risk of electric shock when using the appliance while connected to an electrical outlet.
    • The nebulizer is not protected from liquid penetration. Do not touch the plugged-in device with wet hands or immerse it in water.
    • Do not disconnect the sprayer while the device is running.
    • It is unacceptable to leave a working device unattended.

The compact electronic mesh inhaler WN-114 is intended for treatment and prevention various forms asthma, allergies, colds and viral diseases respiratory system.

New generation technologies!


Information about quality improvements in the B.Well WN-114 nebulizer after replacing the mesh membrane.

Mesh or electronic mesh nebulizer B.Well WN-114– one of the most technologically advanced nebulizers presented on the Russian market. Firstly, it is based on the latest mesh spray technology, which allows the use of an expanded list of drugs for nebulizer therapy. Secondly, this device allows for inhalation with maximum comfort.

What is Mesh? Mesh technology spraying is the “sifting” of drug particles through a mesh membrane. Ultrasonic vibrations of low frequencies are applied not to the drug, but to the membrane mesh, so the molecules of the inhaled substance are not destroyed. The nebulizer allows you to use an expanded range of drugs: antibiotics, mucolytics, hormonal drugs, including Pulmicort and Fluimucil. Mesh technology ensures completely silent operation, high dispersion and spray speed.

The special design of the nebulizer allows for inhalation while holding the device at an angle of up to 45° without compromising the spraying process. This makes the WN-114 convenient for small and even sleeping children and bedridden patients.

B.Well WN-114 is an ergonomic, compact and lightweight device that is convenient to use at home and also easy to take with you.

WN-114 suits the widest range of customers. Due to the possibility of using “serious” drugs, the B.Well mesh nebulizer can be used by patients with chronic obstructive pulmonary diseases (COPD) and asthmatics. The convenience and quietness of the device, as well as the possibility of using mucolytics, makes the WN-114 the first assistant in case of illness in a child. During the period of exacerbations and increased risk of acute respiratory infections - the autumn-winter season - mucolytics will be needed not only for children. For the prevention and treatment of respiratory tract diseases, especially in initial stages or when residual effects, it is safer and more effective not to load the liver with tablets, but to carry out inhalations. It is also suitable for those who want to have a universal device for home and travel.

  • Innovative mesh spray technology
  • Expanded range of inhaled drugs, including Pulmicort and Fluimucil
  • Completely silent operation
  • Possibility of inhalation at an angle of up to 45°
  • Particle size 1.5-4.8 microns
  • Inhalation duration up to 40 minutes
  • Works both from the mains and from AA batteries
  • Convenient storage case
  • Children's and adult masks, mask connector included
  • Ergonomic, compact and lightweight device

Manufacturer: B.Well, UK
Free service maintenance - 10 years
Guarantee- 2 years

Contents of delivery: Inhaler main unit, cap, adult mask, child mask, coupling, mouthpiece, 4 AA batteries, storage bag, user manual, warranty card, packaging.

Operating temperature from +5° to +40°С
Maximum spray capacity 8 ml
Storage temperature from -25° to +60°С
Spray method Low frequency, through a mesh membrane
Dimensions 7.5 x 7.5 x 14.4 cm
Particle size 4.8 microns
Spray speed 0.25 ml/min
Weight 300 g
Operating humidity less than 80%
Power supply 4 x 1.5V AA alkaline batteries
Battery life About 100 min
Ultrasound frequency 100 kHz